Flashlamp replacement for IPL and laser systems
Flashlamps wear out with use, and output falls before the lamp finally fails. We fit a new lamp, check the circuit that fires it, and measure the output once it is fitted.
A new lamp, and proof of what it delivers
A flashlamp is a consumable with a finite number of shots. Replacing the lamp is a repair, and it should not mean replacing the handpiece or the laser head around it.
IPL handpieces
New flashlamps fitted to IPL and SHR handpieces.
Lamp-pumped lasers
Flashlamp replacement on lamp-pumped systems such as Q-switched Nd:YAG lasers.
Firing circuit checked
The capacitor bank that drives the lamp is inspected, since a tired bank shortens the life of a new lamp.
Output measured after fitting
With the new lamp in, output is measured with a calibrated sensor and recorded against the display.
The number on the screen is a setting. We measure what reaches the skin.
Lamps, diode stacks and optics lose output as they age, so a machine can display one figure and deliver another. We measure output at the handpiece with calibrated power and energy sensors, compare it with the display, and give you the readings in writing.
- Measured at the handpiece, not read from the screen
- Calibrated power and energy sensors
- Carried out after every repair
- Recorded at commissioning on every machine we sell
- Written record of the readings for your files
How a repair works
Tell us the fault
Send us the make, model and a description of the fault. We reply with a written quotation, typically within 24 hours.
Collection, drop-off or site visit
We arrange insured collection from your clinic, you drop off at our West Bromwich workshop, or an engineer comes to you.
Diagnosis
We find the failed part. If we find faults beyond the original quotation, we contact you with a revised price before going further.
Repair
The agreed work is carried out at component level, with replaced parts recorded.
Calibration and safety test
Output is measured at the handpiece with calibrated sensors and compared with the display, followed by electrical safety testing.
Return with warranty
Your machine comes back with a service record of the work and the measured output, and a 12-month warranty on the repair.
Tell us about the fault
Send the make, model and what the machine is doing. We reply with a written quotation, typically within 24 hours.